  • The percutaneous lead management kit (PLMK) was developed for the HeartMate 2 (HM2) left ventricular assist device (LVAD) to reduce trauma at the exit site and to maintain a clean environment. REduce Driveline Trauma through StabIlization and Exit Site ManagemenT (RESIST) was a multicenter, prospective, nonrandomized study designed to evaluate the feasibility of the PLMK for managing the HM2 driveline exit site. Fifty patients were enrolled at five sites at a median of 495 days post-HM2 implant; 92% (46 of 50) of patients used the PLMK for a minimum of 30 days. At 30 days, more patients found the PLMK to be extremely comfortable (80% vs. 37%, p < 0.001) and extremely effective at stabilizing the driveline (82% vs. 40%, p < 0.001) compared with each center's standard of care. Frequency of dressing changes was 6-7 days or higher for 85% of the patients with PLMK. Three patients developed driveline infection while on PLMK (6%, 0.15 events per patient year), and 35 patients continued to use the PLMK after 6 months. The PLMK is easy to use, increases patient comfort, and increases driveline stability with a dressing change frequency of 6-7 days.
  • Published In

  • ASAIO Journal  Journal
  • Digital Object Identifier (doi)

    Author List

  • Stahovich M; Sundareswaran KS; Fox S; Hallinan W; Blood P; Chen L; Pamboukian SV; Chinn R; Farrar DJ; Pagani FD
  • Start Page

  • 240
  • End Page

  • 245
  • Volume

  • 62
  • Issue

  • 3